OBJECTIVE: To compare the four-item (PT, APTT, TT, FIB) detection results of hemagglutination reagents (PT, APTT, TT, FIB) in experimental animals SD rats and humans with two different brands of hemagglutination reagents imported and domestic to observe whether there is any difference between them.
METHODS: SPF SD rat and human blood coagulation specimens were selected, and the four indexes of PT, APTT, TT and FIB blood coagulation were detected with two different brands of domestic and imported hemagglutination reagents respectively.
Results: There were differences in the detection results of the four indexes of rat hemagglutination by two different brands of hemagglutination reagents. The test results of rat hemagglutination indexes PT, APTT and FIB domestic reagents were higher than those of imported reagents (P<0.05); The coagulation index of TT imported reagents was higher than that of domestic reagents (P<0.05); however, there was no significant difference in human blood coagulation samples.
Conclusion: There may be differences in the results of the four items of hemagglutination (PT, APTT, TT, FIB) in SD rats detected by different manufacturers and brands of hemagglutination reagents, and each laboratory should establish corresponding background data according to their needs.